Portable Digital Dental Lab X-Ray Machine Imaging System Mobile Unit BLX-8Plus

$ 422.39

Availability: 100 in stock

Description

无标题文档
Description
Product introduction:
This model combines the advantages of similar products found at home and abroad; it eliminates the shortcomings of the on-frequency X-ray machine (high current intensity and excess amounts of scrap x-rays).Using a Toshiba 0.3 micro-focus tube.The tube voltage frequency of this machine is 30 KHz,and the tube current is 1 MA. The radiation scope is at an angle of 24 degrees, located within a distance of 1.2 meters ahead. This is an innovative Chinese design named the "Green X-ray Machine."
TECHNICAL SPECIFICATIONS:
Tube Voltage
60kV
Tube Current
1mA
Time Exposure
1~~9s
High pressure generator
30kHz DC
Rated Power
60w
Focal Spot To Skin Distance
>20CM
Focus Spot Size
0.3mm(TOSHIBA)
Charger Input Voltage
AC110V~220V
Supply Frequency
50Hz±1%
Radiation Leakage
<20ugy/h
Battery
24V DC 8000MA
Weight
4.3KG
Dimensions
37 cm×31 cm×28cm (L×W×H)
Reference of the Angles :
Tooth
Upper
Lower
1-2
42
-15
3
45
-18
4-5
30
-10
6-8
28
-5
Reference of the time of exposure:
Tooth
Adult
Child
Central & Lateral
0.8-1.2
0.8-1.0
Cuspid & 1st Bicuspid
1.2-1.4
1.0- 1.2
Upper 5,6
1.6-1.8
1.4-1.6
Upper 7,8
3.0-3.5
2.8-3.3.
Lower 5,6
1.8-2.0.
1.6-1.8
Lower 7,8
2.0-2.5
1.8-2.3
Packing List:
Main machine 1pc
Power line 1pc
User manual 1pc
Operation of the Main Unit :
Upon re
ceiv
ing,open the box and check the product for possible damage during shipping.
Make sure the fittings on the encasement list are packaged within the box.
Turn on the Power

the pilot lamp will be light up
.
During this time,the digital tube  will show the fore se
ttin
g time. Now the equipment is in standby mode.
Sett
ing the time (Skip to next step if unneeded) Use the "SET","+",and" -" buttons to reset the time as follows: press SET,the pilot lamp above is illuminated, the machine is now in the time setting mode

then press "+" and" -" to set the time needed (time range is 1-9seconds); press set again when the time is adjusted the light lamp is off. The equipment is now in standby mode.
Put the Tooth film plumb behind the tooth which is going to be taken picture, and be as close as possible (the smooth side stick to the t
oot
h)
Keep the ball head plumb to the tooth projection position; Have the ball head,tooth, and tooth film steadily mutually plumb.
After positioning,use the ON/OFF on the main Unit, to take pictures. (Notice: gently press 0.5 sec to turn it on)
After exposure to the tube,press the start button again  in case of an emergency to stop shooting and return to standby status. (At this point there was no output ray,so the dental films can be used again).
Repeat step 6 to reset the picturing time
.
To repeat the picturing process,please repeat steps 7 to 9.
Turn the power  off when the work is
finish
ed,press POWER, making sure the pilot lamp is
off.
Don't keep the ma
chine
long in standby mode.
Points for Attention:
Make sure the angle of film,tooth,and ball head are properly aligned when taking a picture. Keep them steady until the last step of picturing; making sure that there is no change in any of their positions.
Remember to turn the power 0ff when work is done.
Charge the battery if the machine cannot work and the red lamp lignt.
After turning the power on,wait one minute before taking pictures,this allows the ray to provide a steady output.
The equipment switches into protection mode automatically when the voltage is incorrect. If this occurs, it will not be able to carry out the normal work of filming.
When taking pictures,the ON/OFF button could stop the objection and then the machine goes back to the preparation mode.
Use high quality tooth film and developing liquid to make clear pictures.
Handle the ball head gently while in use,so as not to damage the delicate component.
We usually set the time between 3-8 seconds when taking pictures,but the actual time of ray output is 0.2-0.5 seconds.
If a situation occurs, preventing the machine from taking regular pictures, contact the seller to solve the problem instead of trying to service the product yourself.
Keep the tooth film and the liquid in a proper place and use them within the shelf-life period.
Development of films should be performed at temperatures ranging between 23-25 degrees Celsius to ensure the image quality.
Other requests accordance with the scope of the technical parameters.
Please read the manual carefully before using.
FDA Statement
The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, you can bid on this item only if you are an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item.
Seller name:Dentalbox
City,State:SHENZHEN,China
Telephone number:0755-23193580
Device Classification Name
System, X-Ray, Mobile
510(K) Number
K200976
Device Name
SR-8230, SR-8230S Portable X-Ray Unit
Regulation Number
892.172
Classification Product Code
IZL
Date Received
4/13/2020
Decision Date
6/10/2020
Decision
Substantially Equivalent (SESE)
Regulation Medical Specialty
Radiology
510k Review Panel
Radiology
Summary
Summary
Type
Traditional
Reviewed By Third Party
No
Combination Product
No
Relevant Policy